Test Modes & Standards
- NSS: Neutral Salt Spray Test.
- AASS: Acetic Acid Salt Spray Test.
- Conformity Standards: ASTM B117, ISO 9227, JIS Z2371, DIN 50021
Hardware Highlights
- Box material: Uses P.V.C or P.P sheets (acid and alkali resistant, anti-aging, high-temperature resistant).
- Precision nozzle: Utilizes specially designed glass nozzles (no crystalline blockage, uniform spray diffusion).
- Top cover design: V-shaped or arc-shaped top cover (prevents condensate water from dripping onto the test sample).
- Sealing method: Water-sealed structure.
- Heating method: Steam heating method, with fast heating and uniform distribution.
Model
| TA-SS | ||||||
60 | 90 | 120 | 160 | ||||
Lab temp | Salt water test 35℃±1℃,corrosion resistance test 50℃±1℃ | ||||||
Saturation bucket temp | Salt water test 47℃±1℃,corrosion resistance test 63℃±1℃ | ||||||
PH value | Salt water test 6.5~7.2,corrosion resistance test 3.0~3.2 | ||||||
Temp deviation | ±2℃ | ||||||
Inner size(W*H*Dcm)
| 60x40x45 | 90x50x60 | 120x50x100 | 160x50x100 | |||
Outer size(W*H*Dcm)
| 113x107x63 | 146x128x91 | 190x140x120 | 224x150x160 | |||
Water tank capacity (L) | 15 | 25 | 40 | 60 | |||
Spray amount (ml/h) | 1.0~2.0 | ||||||
Mist removal method | Air compression | ||||||
Protector | Water level controller, 3 status warning lights, 2 sets of over temperature protection switches | ||||||
Standard accessories | One set of storage sticks, 500g bottles of NaCl reagent, one saline mixing bucket, one 80cm ² standard collection cup, one 50ml measuring cylinder, and one detailed operating manual | ||||||
Can the salt spray testing machine conduct the cyclic corrosion test (CCT)?
How to calibrate the salt spray deposition rate?
- Tools for preparation: Use a standard funnel with a diameter of 100mm (with a collection area of approximately 80cm²) and connect it to a graduated cylinder.
- Placement: Place at least two collectors in the box: one close to the nozzle and one far from the nozzle. Ensure there is no sample obstruction above and no condensate dripping from the top cover is collected.
- Test duration: Conduct continuous spray tests for at least 16 hours (minimum of no less than 8 hours) to obtain accurate average values.
- Criteria for judgment: The volume of salt solution collected per hour should be between 1.0 and 2.0 ml / 80cm² / h.
- How to adjust: If the sedimentation rate is too low: Lower the top of the spray tower (diffusion cone) or slightly increase the intake pressure.
What are the differences between the salt used for testing and industrial salt?
- Purity requirement: The standard salt spray test (such as ASTM B117) stipulates that the purity of sodium chloride (NaCl) must be ≥ 99.5%.
- Issues with table salt: Contains iodine (which interferes with the chemical reaction and affects the test accuracy) and anti-caking agents (which can clog the nozzle).
- Problems with common industrial salt: Contains heavy metal impurities (such as copper and nickel), which accelerate corrosion and cause the test results to deviate from the standard.